Tourniquet Overview
PREFACE
To the layperson, a tourniquet is a cloth - and - stick device
used to stop bleeding in an emergency. Yes, that is a tourniquet.
But modern surgical tourniquets are also much more. Today,
surgical tourniquets are specifically designed to enable surgeons
to perform delicate dissections in a bloodless operative field.
They use compressed gas to apply a carefully controlled amount of
pressure to an extremity. Some computerized tourniquet systems
perform self - checks of calibration, display elapsed inflation
time, and sound alarms if problems arise. And problems can arise:
equipment can malfunction and patients can be injured. The
perioperative nurse shares the responsibility for protecting
patients from hazards related to tourniquet use. Surgery is a
frightening experience for many patients; ideally, they can be
spared the additional anguish of nerve injury, compartment
syndrome, prolonged swelling, toxic reactions, and other
tourniquet - related complications.
Historical
Perspective
Compression dressings to control bleeding are described in
papers attributed to ancient Greeks from the renowned medical
school at Cos. As far back as Roman times, military surgeons used
compressing devices to control bleeding during amputations
| In 1628, William Harvey, an English surgeon, paved
the way for future technological developments by tracing
the human circulation. In 1718, Louis Petit, a French
surgeon, developed the screw device shown in Figure 1.
From the French verb "tourner" (to turn), he
named the device "tourniquet." Joseph Lister
(1827 - 1912) is credited with being the first to use a
tourniquet to create a bloodless surgical field in 1864.
For exsanguination, he recommended elevation of a limb
for 4 minutes before applying the tourniquet. In 1873,
Johann von Esmarch devised a rubber bandage for
exsanguination and tourniquet use. The device was
superior to Petit's screw device, because Petit's cloth
bandages tore and the screw could untwist. In 1881,
Volkmann demonstrated that limb paralysis could result
from use of the Esmarch tourniquet.
|
 |
| Figure 1. Louis Petit's tourniquet,
1718. |
|
An inflatable (pneumatic) tourniquet was developed by Harvey
Cushing in 1904. To constrict the blood vessels, compressed gas
was introduced into a cylindrical bladder. This device had two
advantages over the Esmarch tourniquet: (1) rapid application and
removal; and (2) decreased incidence of nerve paralysis.
The use of two tourniquets for administering segmental
anesthesia was introduced by August Bier in 1908. In this two -
tourniquet procedure, circulation is isolated in a portion of the
extremity and anesthesia is infused intravenously. The procedure
did not become popular at that time; however, in 1963, Holmes
reintroduced it using the single - tourniquet technique and
improved anesthesia. Today, the two - tourniquet technique is
used frequently. This so - called intra - venous regional
anesthesia (IVRA) is commonly referred to as a Bier Block, or
Bier's method.
In the early 1980's, modern electronic tourniquet systems
(also called computerized tourniquets or microprocessor -
controlled tourniquets) were invented by James McEwen, PhD, a
biomedical engineer in Vancouver, Canada. The first US patent for
such a modern electronic tourniquet system was awarded to Dr.
McEwen in 1984, and to date he has been awarded many more US and
foreign patents for tourniquet improvements. The introduction and
use of automatic tourniquet systems based on Dr. McEwen's
inventions has significantly improved safety and convenience over
previous mechanical pressure regulator systems. Most of these new
tourniquet systems are self - calibrating and self - contained
(not requiring an external high pressure gas source) and provide
a variety of safety features not possible in older mechanical
tourniquets.
Definition of
Tourniquets
A tourniquet can be defined as a constricting or compressing
device used to control venous and arterial circulation to an
extremity for a period of time. Pressure is applied
circumferentially upon the skin and underlying tissues of a limb;
this pressure is transferred to the walls of vessels, causing
them to become temporarily occluded. In surgical settings, a
tourniquet is used following exsanguination to produce a
relatively bloodless operative field.
Types of Tourniquets
Two distinct types of tourniquets are found in the surgical
setting:
- Noninflatable (nonpneumatic) tourniquets constructed of
rubber or elasticized cloth.
- Pneumatic tourniquets, which have cuffs that are inflated
by compressed gas.
The surgical use of noninflatable tourniquets is very limited.
Historically, the use of a surgical glove as a wrist tourniquet
for hand surgery has been reported, as has the use of a Penrose
drain for digit surgery. However, in surgery today, noninflatable
tourniquets have largely been supplanted by the safer and more
convenient use of modern electronic tourniquet systems connected
to inflatable cuffs. For phlebotomy or intravenous infusion,
simple rubber tubing may be utilized. Elastic bandages are used
to control bleeding following procedures such as vein stripping.
Plain cloth bandages and dowel tourniquets are used primarily in
nonsurgical settings. For prehospital care of a patient with
trauma to an extremity, a nonpneumatic tourniquet may be employed
as a last resort to control hemorrhage.
A pneumatic tourniquet uses a gas - inflated cuff to constrict
blood flow. A regulating device on the tourniquet apparatus can
be preset to control the amount of cuff pressure exerted on the
limb.
Components
of Pneumatic Tourniquets
Modern pneumatic tourniquets have five basic
components:
- An Inflatable Cuff.
- A Compressed Gas Source.
- A Pressure Display.
- A Pressure Regulator.
- Connection Tubing.
Inflatable Cuff
Pressure is exerted on the circumference of an
extremity by means of compressed gas, which is introduced
into a bladder within the tourniquet cuff. All bladders
have one or two port connectors for the attachment of
connecting tubing. Typically, pneumatic cuffs are
fastened by contact closures and may be secured with a
ribbon tie to prevent cuff movement during the procedure.
|
 |
| Figure 2. Modern tourniquet cuffs:
Reusable single bladder standard rectangular cuff
(top left), reusable contour lower leg cuff
(center), sterile disposable single bladder cuff
(lower right), and reusable pediatric cuff
(bottom center). |
|
Many different types of cuffs are available, and the
appropriate choice is determined primarily by proper fit and
surgical procedures. The choice of a tourniquet cuff should be
individualized, taking into consideration the size and shape of
the patient's limb and the specific demands of the operative
procedure. When selecting a cuff, consider the following
criteria:
- Cuff Location.
- Single - vs. Dual - Bladder Design.
- Cuff Shape.
- Cuff Length.
- Cuff Width.
- Disposable vs. Reusable Cuffs.
- Specialty Applications.
- Limb Protection.
| Cuff Location Different
cuffs are designed to be placed on different extremities.
For most lower extremity procedures, select a thigh cuff.
For foot procedures where it is desirable to place the
cuff above the ankle on the lower third of the calf
select a lower leg cuff. Most upper extremity procedures
call for a standard cuff applied to the upper arm,
although hand surgeons may utilize a small tourniquet
inflated around the wrist, or even two relatively narrow
pneumatic cuffs on the forearm.
Single
vs. Dual - Bladder Design
The surgical procedure will also influence the
surgeon's choice of a single - bladder cuff or a dual -
bladder cuff (one with two narrow bladders within a
single cuff, see Figure 3). For general anesthesia cases,
such as total knee arthroplasty, a single - bladder cuff
is usually used. For intravenous regional anesthesia (or
IVRA, also known as Bier block anesthesia), a dual -
bladder cuff is usually applied. Inflation and deflation
of each bladder of a dual - bladder cuff is controlled
separately; this permits greater safety and patient
comfort, particularly for longer procedures.
Cuff Shape
Standard rectangular (or cylindrical) tourniquet cuffs
are designed to fit optimally on cylindrically shaped
limbs. However, human limbs may be conical in shape
(particularly in extremely muscular or obese individuals)
which can result in poor fit, sliding of the cuff
distally on the limb during the procedure, and inability
to achieve a bloodless field at normal pressures if
standard rectangular cuffs are used. Contour cuffs have
an arced design that gives them a smaller diameter
distally than proximally when wrapped (see Figure 4).
Some contour cuffs also have pivoting fastening straps
that allow the proximal and distal diameters to be
adjusted to suit the limb. Contour cuffs enhance comfort
in patients with conically shaped limbs and reduce the
risk of mechanical shearing. It has been reported that
contour tourniquets occlude blood flow at lower inflation
pressure than standard rectangular cuffs of equal width,
which may be attributable to better cuff fit and more
efficient transmission of pressure to deep tissues.
Cuff Length
Perioperative nurses often assume responsibility for
selecting an appropriate cuff size. A cuff that is too
long or too short can cause problems.
Selecting a cuff that is too long (resulting in
excessive overlap as shown in Figure 5 may be difficult
to apply snugly and may be less stable (causing the cuff
to move distally or "walk" on the patient's
limb). Both problems may prevent occlusion of arterial
bloodflow at normal cuff pressures, lead to loss of
occlusion during the procedure, and lead to skin injury.
Selecting a cuff that is too short (resulting in too
little overlap of the inflatable bladder portion of the
cuff as shown in Figure 6) produces uneven distribution
of pressure and can lead to loosening of the cuff or an
inability to sustain occlusion.
On some cuffs the inflatable bladder portion extends
the full length of the cuff, while on other types of cuff
there is a 1 - 2 inch long non - inflating portion at one
end through which the strap is sewn.
Some manufacturers color code their cuffs to assist
the user in selecting the most appropriate cuff for the
patient. Note that these color codes may differ between
the various cuff manufacturers.
To determine the appropriate cuff length, measure the
circumference of the limb near the middle of the location
chosen for the cuff and refer to the cuff manufacturer's
instructions for the appropriate cuff size.
In general, the shortest cuff giving at least the
manufacturer's specified minimum overlap should be used.
Some cuffs have an additional hook - type fastener patch
on the inside surface near the strap end for additional
safety and to assist with application of the cuff to the
limb. If this patch does not fully engage the loop
fastener on the outside of the cuff, overlap is
insufficient and a longer cuff must be selected (see
Figure 7). Maximum overlap varies depending on cuff
manufacturer, cuff type, and limb size but is typically
between ¼ and ½ of the overall cuff length.
|
 |
| Figure 3. Dual bladder IVRA cuff
(left) and single bladder cuff (right), both dual
port. |
| |
 |
| Figure 4. Standard cylindrical cuff
(left) wide contoured lower leg cuff (right) |
| |
 |
| Figure 5. Excessive cuff overlap
condition. |
| |
 |
| Figure 6. Non - overlapping condition
(gap between cuff ends). |
|
| |
 |
| Figure 7. Dual Fastening Feature |
|
Some cuffs are supplied in a single size for each
limb location (e.g. adult arm, thigh, or lower leg) and
are designed with a range sufficient to fit 95% of the
patient population. With these cuffs ensure that the limb
circumference falls within the range specified in the
manufacturer's cuff instructions and that the fasteners
engage properly as outlined above. Pediatric patients or
adults with unusually small limbs require extra care in
tourniquet practice and normally require specially
designed cuffs and lower inflation pressures. |
Table 1. Single bladder Tourniquet Cuffs
| Recommended Cuff Size |
Zimmer ATS Cylindrical Cuff |
Zimmer Sterile Disposable Cuff |
| |
Limb Circumference Range |
Limb Circumference Range |
| 8 in. (20 cm) |
6.0-7.5 in. (15-19 cm) |
4.0-6.5 in. (10-17 cm) |
| 12 in. (30 cm) |
7.5-11.0 in. (19-28 cm) |
6.5-10.0 in. (17-25 cm) |
| 18 in. (46 cm) |
10.5-16.5 in.(27-42 cm) |
10.0-17.0 in.(25-43 cm) |
| 24 in. (61 cm) |
16.5-23.0 in. (42-58 cm) |
17.0-23.5 in. (43-60 cm) |
| 30 in. (76 cm) |
21.0-28.0 in. (53-71 cm) |
23.5-29.5 in. (60-75 cm) |
| 34 in. (86 cm) |
22.5-32.0 in. (57-81 cm) |
29.5-33.5 in. (75-85 cm) |
| 42 in (107 cm) |
28.0-40.0 in. (71-102 cm) |
33.5-41.5 in. (85-105 cm) |
Individual manufacturers may offer different sizes than those
listed in Table 1. The 8 - inch (20 - cm) cuff is intended for
thin or small limbs. Generally, it is used with lower inflation
pressures than the other cuff sizes.
Table 2. Dual Bladder Tourniquet Cuffs
| Recommended Cuff Size |
Zimmer ATS Cylindrical Cuff |
Zimmer Sterile Disposable Cuff |
| |
Limb Circumference Range |
Limb Circumference Range |
| 12 in. (30 cm) |
7.5-11.0 in (19-28 cm) |
6.5-10.0 in. (17-25 cm) |
| 18 in. (46 cm) |
10.5-16.5 in. (27-42 cm) |
10.0-17.0 in. (25-43 cm) |
| 24 in. (61 cm) |
16.5-23.0 in. (42-58 cm) |
17.0-23.5 in. (43-60 cm) |
Cuff Width
Select as wide a cuff as possible. It has been shown that a
cuff with a wider bladder occludes blood flow at a lower pressure
level than does a cuff with a narrow bladder. This may be related
to more efficient pressure transmission to the deeper tissues
with a wider cuff. The lower pressure may reduce the risk of
pressure related injury to the patient. Extra care must be taken
with unusually small adult patients and pediatric patients to
ensure that the correct cuff width is used and the cuff edges do
not lie close to the joints of the limb to reduce the risk of
nerve injury.
Disposable vs.
Reusable Cuffs
Sterile, disposable cuffs are available for situations that
require placement of a sterile tourniquet near the operative
site, or for use in contaminated surgical cases. The design and
materials of disposable cuffs are suitable for a single
sterilization cycle and single use only and must not be
resterilized or re - used. If a disposable cuff is selected, it
must be discarded at the end of the procedure. Reusable cuffs are
not designed to be sterilized and must be used with sufficient
sterile draping to isolate the cuff from the sterile field.
Specialty
Applications
Patients with unusually small or large limbs may require
specialized tourniquet cuffs. Some specially designed pediatric
cuffs have optimized width appropriate for short limb segments
and thinner, more supple construction suitable for wrapping
around small diameter limbs. Fitting the cuff and maintaining
occlusion of arterial bloodflow may be difficult with obese
patients and particular care must be taken. Standard cuffs may
not be large enough in some cases, and custom cuffs may be
required.
Limb Protection Beneath
Cuffs
For some cuffs, a matching limb protection sleeve is available
to help protect the soft tissues under the cuff. These sleeves
are sized to give optimal protection to limbs within the
recommended size range of the matching cuff and can help reduce
wrinkling, pinching and shearing of the soft tissues. A sleeve
that is not specifically matched to the cuff being used may not
protect the underlying soft tissues and may impair the
performance of the cuff. Some tourniquet cuff manufacturers color
code their matching limb protection sleeves to help ensure that
the correct sleeve is used with the selected cuff. As this color
coding is specific to each manufacturer, tourniquet cuffs and
sleeves from different manufacturers should never be mixed.
Figure 8 shows a comparison of molds taken under a typical
tourniquet cuff with no underlying limb protection, with only
cast padding used as limb protection, and with a limb protection
sleeve specifically matched to the tourniquet cuff being used.
 |
 |
 |
| (a) |
(b) |
(c) |
| Figure 8. Comparison of molds
taken under a typical cuff with: a) no limb protection
sleeves b) cast padding c) a limb protection sleeve
matched to the cuff being used. |
Tourniquet
Instruments
| Compressed Gas Source The tourniquet cuff
bladder requires a source of compressed gas to supply a
carefully controlled amount of tourniquet pressure. The
gas used may be nitrogen, ambient air, or some other gas.
Some tourniquet systems utilize high - pressure gas,
while other systems use low - pressure gas. Never use
nitrous oxide or oxygen to inflate the tourniquet cuff,
because of the increased risk of fire.
Some modern electronic tourniquet systems (also called
microprocessor - controlled or computerized tourniquet
systems) utilize an internal electrical pump to compress
the ambient air; these systems do not require an external
pressure source. Others are designed to use external
pressure sources, such as portable canisters, portable
tanks, or built - in hospital systems.
Pressure Display
The pressure display is a device that visually
indicates the amount of pressure in the tourniquet cuff
bladder. All pneumatic tourniquets have a pressure
display; in older non - electronic systems it is normally
a dial (aneroid) gauge. In most modern electronic
systems, pressure is shown on a microprocessor -
controlled digital display.
|
 |
| Figure 9. Dual port tourniquet
instrument with two channels for two separate
cuffs or bladders |
|
Pressure Regulator
The pressure regulator adjusts and controls the gas pressure
in the cuff bladder. Non - computerized tourniquet systems
utilize valves that respond mechanically to changes in pressure.
For example, if pressure in the cuff bladder falls, a valve opens
to allow more gas to enter the regulator from the gas source; if
pressure exceeds a certain level, the pressure forces a release
valve to open and expel gas into the environment. Sometimes, the
pressure levels at which these two valves turn on and off are
quite different and cuff pressure may fluctuate within a certain
range above and below the selected pressure. Due to the sensitive
mechanical components of these systems, it is very important to
follow the manufacturer's instructions regarding testing and
calibration and to perform these checks before each surgical
procedure as recommended.
In most modern electronic tourniquet systems, the internal
electrical pump, pressure display, and pressure regulator are
combined in a single instrument in which a microprocessor
continuously monitors and compensates for changing levels of
pressure in the cuff bladder. Regulation does not rely on
mechanical (pressure) forces to turn valves off and on. Instead,
the microprocessor can detect extremely small changes in the cuff
pressure and regulate the inflow of gas to control the pressure.
Some modern electronic tourniquet systems use a sophisticated
"dual port" system which gives the most accurate
control of cuff pressure and the fastest response to pressure
changes. In a dual port system, each bladder has two ports and a
double cuff - to - instrument hose. One port is for monitoring
the pressure ("sensing port"); the other port is for
inflating and deflating the cuff and for automatically supplying
and releasing small amounts of gas during use to control cuff
pressure ("supply port"). In some more basic systems a
single port performs both functions for each cuff.
The tourniquet instrument may provide one or two channels,
allowing one or two cuffs to be used simultaneously. Figure 9
shows a dual port two channel tourniquet instrument. A dual cuff
control valve is sometimes added to a single channel system to
allow use of two separate cuffs or bladders, typically for IVRA
(Bier block) procedures.
Modern electronic tourniquet systems include many safety
features to help improve patient safety when working with a
pneumatic tourniquet. While some tourniquet systems may provide
much in the way of sophisticated safety monitoring and
interlocks, perioperative personnel should be aware of
potentially hazardous conditions and monitor the tourniquet
system during the time the cuff is applied to the patient.
Potential hazards include the following:
- Kinked or occluded tubing connecting the cuff to the
instrument may prevent the instrument from displaying the
correct cuff pressure, or controlling the cuff pressure
correctly. Some modern electronic tourniquet instruments
include sophisticated monitoring techniques to warn users
of tubing occlusions.
- Inadvertent deflation of both cuffs during a Bier Block
(IVRA) procedure due to operator error may allow a bolus
of anesthetic to enter the circulatory system suddenly
and prematurely. Some modern electronic tourniquet
instruments include special safety interlocks
specifically to help prevent such inadvertent deflation
of both cuffs.
- Leaking cuffs or leaking tubing connections may prevent
the tourniquet instrument from maintaining the set
pressure in the cuff. Alarms are provided in some modern
tourniquet instruments to warn of leaking cuffs or low
cuff pressures.
- If the power switch of an some types of electronic
tourniquet instruments is inadvertently switched to an
'off' or 'standby' position while the cuff is still
pressurized (due to operator error), then the pressure
will be maintained in the cuff to maintain patient safety
but the pressure display, timer and alarms of the
tourniquet instruments may no longer function and so the
operator may no longer be aware of the cuff pressure or
tourniquet time. Some of the most modern tourniquet
instruments include safety interlocks to prevent the
power from being switched to 'off' or 'standby' positions
while a pressurized cuff is connected.
The use of adaptors or accessories not approved by the
instrument manufacturer may interfere with the ability of some
types of modern electronic tourniquet instruments to detect alarm
conditions and potentially hazardous conditions. Read the
instruction manual for the specific tourniquet instrument(s) in
your facility and be sure you understand the alarms and safety
features of your specific tourniquet instruments.
| Connecting Tubing Most pneumatic
tourniquet systems use a hose assembly between the
tourniquet instrument and the cuff utilizing newer
Positive Locking Connectors or older Luer - lock
connectors (see Figure 10). Some pneumatic tourniquets
require an additional hose assembly between the external
compressed gas source and the regulator.
Note that additional care must be taken with the older
twist - type Luer - lock connectors because they may
accidentally disconnect if the hoses are twisted during
movement of the hoses, cuff, patient, or instrument
during use. Newer Positive Locking Connectors are less
likely to disconnect accidentally.
Additional Features
Manufacturers of some pneumatic tourniquets have a
calibration kit available for checking the tourniquet
regulator and pressure display. Some modern electronic
tourniquet systems perform a self - calibration each time
the power is switched on.
Additional features of modern electronic systems,
often built into the regulating equipment, are alarm
systems to detect unusual increases or decreases in
pressure, to indicate elapsed inflation time and alert
staff when a pre - selected maximum tourniquet time has
been exceeded, and to warn of a possible kinked or
blocked hose, disconnected hose, or leak in the system.
|
 |
| Figure 10. Positive Locking
connectors (top and middle), and older style Luer
Lock connectors (bottom) |
|
To help prevent accidental release of anesthetic in dual cuff
Bier Block (IVRA) procedures, some systems include a warning
function requiring the user to confirm deflation of the last of
the two cuffs before that cuff is actually deflated. Batteries
are included in some electrically operated tourniquets, enabling
a patient to be transported with the cuff(s) inflated, if
necessary, and so that the tourniquet will continue to function
in the event of an electrical failure. Such systems usually
include an indicator or warning if battery power becomes low and
recharging is required. The most advanced of the computerized
tourniquet systems also incorporate safety functions to warn the
user and prevent the electrical power to the system from being
accidentally switched to 'off' or 'standby' positions if the user
makes an error and attempts to do so while a tourniquet cuff is
still inflated.
|